AZ8601
View model →
AZ8403
Handheld dissolved oxygen meter for aquaculture, water treatment, lab DO tracking. Clark-type DO probe, salinity & barometric compensation, 99-point memory, USB/IrDA output.
Reference price: USD 146.00
For optimal accuracy, calibrate before each use or daily. Recalibrate after replacing the membrane or when there are significant changes in temperature or altitude.
Replace them if you notice unstable readings, slow response, or difficulty calibrating. Regular maintenance is recommended every 3–6 months, or immediately if the membrane is torn, fouled, or dry.
Yes, but please note that while the AZ 8403 provides high-accuracy DO readings, it does not have a built-in BOD calculation mode. Users must design their own testing protocol to manually calculate the BOD value based on the results.
It uses the most cost-effective Clark type electrode. The 8403 DO probe does not need to refill electrolyte often as a galvanic electrode.
The 8403 stores and recalls 99 data points with temperature readings, each with a real-time clock timestamp.
USB interface and IrDA (infrared) interface. Data can be transmitted to the 9680 IrDA printer or linked to a PC via USB cable.
Yes, both can affect DO accuracy. The 8403 allows manual salinity adjustment (0.0–50.0 ppt) and barometric pressure adjustment (500–1499 mmHg) for compensation.
Over 6 months with good maintenance.
Specifications
| DO Range | 0–19.99 mg/L (ppm) |
|---|---|
| DO Accuracy | ±1.5 % FS |
| Temperature Range | 0–50 °C |
| Temp accuracy | ±0.3 °C |
| Salinity comp | 0.0–50.0 ppt |
| Baro comp | 500–1499 mmHg |
| Memory | 99 pts |
| Sensor | Clark cell . |
| Power | 4×AAA . |
| Weight | ~150 g |